| Customization: | Available |
|---|---|
| Classification: | Imaging Diagnostic Equipment |
| Type: | ECG |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features of handheld ecg equipment:
|
1. 8" color TFT LCD screen, resolution 800*600 |
|
2. Waterproof design, resistance to dropping (1 meter) |
|
3. Low power consumption, fanless structure which creates no noise |
|
4. Support external 13-20V DC power, continuously working for 5.5 hours above on battery backup, full charge with 2. 5 hours, suitable for Ambulance, Emergency Transport |
|
5. Display mode: Big font, trends review, 7 leads ECG waveform on same screen, and etc |
|
6. 15 kinds of arrhythmia analysis, 15 kinds of drug concentration analysis |
|
7. Anti-electrosurgical unit (300W), anti-defibrillation (360J), suitable for Operating Room, General Ward |
|
8. Digital SpO2 technology, excellent anti- interference performance, low perfusion (0.1%) |
ECG
|
Lead: Standard 3/5 lead |
|
Lead Mode: 5 Leads (RA, LA, LL, RL,V) |
|
Lead Selection: I, II, III, avR, avL, avF, V |
|
Lead Mode: 3 Leads (RA, LA, LL) |
|
Lead Selection: I, II, III, |
|
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV |
|
HR and Alarm Range: Adult: 15 ~300bpm Ped/Neo: 15 ~ 350 bpm |
|
Accuracy: ±1% or±1bpm, which is greater |
|
Resolution: 1 bpm |
|
Sensitivity: > 200 (uVp-p) |
|
Differential Input Impedance: > 5 MΩ |
|
Bandwidth: Surgery: 1 ~ 20 Hz Monitor : 0.5 ~ 40 Hz Diagnostic: 0.05 ~ 130 Hz |
|
CMRR: Monitor > 100 dB Operation > 100 dB Diagnosis > 90 dB |
|
Electrode Offset Potential: ±300mV |
|
Range: ±2 mV ~ ±700 mV |
|
Width: 0.1 ms ~ 2 ms |
|
Rise time: 10 us ~ 100 us |
|
ST Segment Monitoring Range: Measurement Range: -2.0 ~ +2.0 mV |
|
Measurement Accuracy: -0.8 ~ +0.8 mV |
|
Measurement Error: ±0.02 mV or ±10% |
RESP
| Method: Impedance Between RA-LL or RA-LA |
| Measuring Impedance Range: 0.3~3 Ω |
| Base line Impedance Range: 200~4 KΩ |
|
Bandwidth: 0.1~2.5Hz |
| Resp. Rate: |
| Measuring and Alarm Range: |
| Adult: 2 ~ 120 BrPM |
| Neo/Ped: 2 ~ 150 BrPM |
| Resolution: 1 BrPM |
| Accuracy: ±2 BrPM |
| Apnea Alarm: 10 ~ 40 S |
NIBP
|
Method: Oscillometric |
|
Mode: Manual, Auto, STAT |
|
Measuring Interval in Auto Mode: |
|
1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min) |
|
Measuring Period in STAT Mode: 5 Min |
|
Pulse Rate Range: 40 ~ 240 bpm |
|
Measuring and Alarm Range: |
|
Adult Mode |
|
SYS 40 ~ 270 mmHg |
|
DIA 10 ~ 215 mmHg |
SpO2
|
Measuring Range: 0 ~ 100 % |
|
Resolution: 1 % |
|
Accuracy: 70% ~ 100% ±2 % 0% ~ 69% unspecified |
|
Pulse Rate: Measuring Range: 20~300 bpm Resolution: 1bpm Accuracy: ±3bpm |
TEMP
|
Applicable Temperature Sensor: YSI, CYF |
|
Channel: 1 |
|
Measuring Range: 0 ~ 50 ºC |
|
Resolution: ±0.1 ºC |
|
Accuracy: ±0.1 ºC (Do not including sensor error) |
|
Dimension: 210*80*175mm (L*W*H) |
|
Weight: 1.35kg |
|
Standard: 3/5-Lead ECG, RESP, SpO2, NIBP, PR, 1-TEMP |
|
Optional: Mainstream/Sidestream EtCO2, Touch screen |