Customization: | Available |
---|---|
Type: | Detecting Vital Signs from Animals |
Product Name: | Veterinary Monitor |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Standard Configuration
|
Option Configuration
|
|
Main machine
|
1 PC
|
ETCO2, Dual IBP, SPO2
|
ECG lead wire
|
1 PC
|
Suntech NIBP, Anesthetic gas module, Printer
|
SPO2 sensor
|
1 PC
|
Touch screen function
|
BP cuff
|
VET with different size
|
Trolley, Wall mount
|
SPO2 extension cord
|
1 PC
|
Center monitoring system
|
BP extension cord
|
1 PC
|
12 lead ECG
|
Power cord
|
1 PC
|
|
Temperature probe
|
1 PC
|
item
|
value
|
Place of Origin
|
China
|
Brand Name
|
ICEN
|
Model Number
|
IN-C100
|
Properties
|
Detecting Vital Signs from Animals
|
Product name
|
Veterinary Monitor
|
Screen
|
12.1 inches color Touch screen
|
Three working modes
|
Monitoring, surgery and diagnosis
|
Output
|
VGA for external display
|
Standard 6 parameters configuration
|
SPO2, ECG, TEMP, PR/HR, NIBP, RESP
|
Equipment size
|
305.6mm*150mm*267mm
|
ECG
|
SPO2
|
NIBP
|
Lead Mode: 5 Leads (I, II, III, AVR, AVL,AVF, V)
|
Measuring Range: 0 ~ 100 %
|
Method: Oscillometry
|
Gain: 2.5mm/mV, ´5.0mm/mV, 10mm/mV, ´20mm/mV
|
Resolution: 1%
|
Measure mode: Manual, Auto, STAT
|
Heart Rate: 15-300 BPM (Adult); 15-350 BPM (Neonatal)
Resolution: 1 BPM Accuracy: ±1% Sensitivity >200 uV(Peak to peak) |
Accuracy: 70% ~ 100% (±2 %)
|
Unit: mmHg, kPa
|
ST measurement range: -2.0 ~ +2.0 mV
Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater Other Range: unspecified |
Pulse Rate: 20-250 BPM
|
Measure and alarm range:
Adult Mode: *SYS 40 ~ 280 mmHg * DIA 10 ~ 225 mmHg * MEAN 20 ~ 240 mmHg Pediatric Mode: * SYS 40 ~ 220 mmHg * DIA 10 ~ 160 mmHg * MEAN 20 ~ 170 mmHg Neonatal Mode: * SYS 40 ~ 135 mmHg * DIA 10 ~ 100 mmHg * MEAN 20 ~ 110 mmHg |
Bandwidth:
* Diagnostic: 0.05~130 Hz * Monitor: 0.5~40 Hz * Surgery: 1~20 Hz |
Resolution: 1 BPM
|
Resolution: 1mmHg
|
|
Accuracy: ±3 BPM
|
Accuracy: ±5mmHg
|